Product Manual

RescueCoach™ Voice and Text Prompts
Table A-8: CPR feedback device (optional)
Text display Prompt level
Voice prompt
Line 1
Line 2 Situation Adv Std Bas
Remove green square package
from lid of AED.
REMOVE GREEN SQUARE
PACKAGE FROM AED LID
Prompts at the beginning of a CPR session. The green
package contains the CPR device.
X X
Tear open green package and
remove CPR device.
TEAR OPEN PACKAGE
REMOVE CPR DEVICE
Prompts to remove the CPR device. X X
Place CPR device on center
of patient’s chest, between
nipples
PLACE DEVICE ON
CENTER OF CHEST
Prompts rescuer to put the CPR device in the correct
place for giving compressions.
X X
Place heel of one hand on CPR
device.
PLACE ONE HAND ON
CPR DEVICE
Prompts rescuer to place one hand on the CPR
device.
X X
Press Slower PRESS SLOWER If the rescuer is giving compressions too fast, prompts
to slow the rate.
X X X
Press Faster PRESS FASTER If the rescuer is giving compressions too slowly,
prompts to quicken the rate.
X X X
Press Softer PRESS SOFTER If the rescuer is giving compressions that are too
deep, prompts to lessen the depth.
X X X
Press Harder and fully release PRESS HARDER
FULLY RELEASE
If the rescuer is giving compressions that are too
shallow, prompts to use more eort and release all
pressure when moving hands up.
X X X
A-6 Powerheart® G5 Automated External Debrillator70-01704-01 B