Product Manual
RescueCoach™ Voice and Text Prompts
Table A-5: Delivering therapy - G5 semi-automatic (continued)
Text display Prompt level
Voice prompt
Line 1
Line 2 Situation Adv Std Bas
Shock not delivered. SHOCK NOT DELIVERED Plays for either of these situations:
• The SHOCK button is not pressed within 30
seconds of the AED giving the “Press red ashing
button...” prompt.
• The AED is unable to deliver a shock because of a
fault condition.
X X X
It is now safe to touch the
patient.
NOW SAFE
TO TOUCH PATIENT
Advises the rescuer that it is safe to touch the patient:
• After the AED delivers a shock
• After the AED detects a nonshockable rhythm
X X
Table A-6: Delivering therapy - G5 automatic
Text display Prompt level
Voice prompt
Line 1
Line 2 Situation Adv Std Bas
Shock will be delivered in SHOCK IN: Noties the rescuer after the AED is fully charged and
ready to deliver the shock.
X X X
Three THREE Prompts approximately three seconds prior to
delivering shock.
X X X
Two TWO Prompts approximately two seconds prior to
delivering shock.
X X X
One ONE Prompts approximately one second prior to
delivering shock.
X X X
Shock delivered. SHOCK DELIVERED Prompts when the shock is delivered. X X X
Shock not delivered. SHOCK NOT DELIVERED Plays if the AED is unable to deliver a shock because
of a fault condition.
X X X
It is now safe to touch
the patient.
NOW SAFE
TO TOUCH PATIENT
Advises the rescuer that it is safe to touch the patient:
• After the AED delivers a shock
• After the AED detects a nonshockable rhythm
X X
A-4 Powerheart® G5 Automated External Debrillator70-01704-01 B