Product Manual
RescueCoach™ Voice and Text Prompts
Table A-3: Pads prompts
Text display Prompt level
Voice prompt
Line 1
Line 2 Situation Adv Std Bas
Pediatric Pads Connected PEDIATRIC
PADS
Noties rescuer that the pediatric pads are connected
to the AED.
X X X
Make sure pad connector is
plugged into AED
CHECK CONNECTOR
IS PLUGGED INTO AED
Prompts when debrillation pads connector is not
inserted into the pad socket.
X X X
Press pads rmly to patient’s
bare chest
PRESS PADS FIRMLY
TO CHEST
Prompts when better pad contact to the patient’s
skin is required.
X X X
Table A-4: Analysis
Text display Prompt level
Voice prompt
Line 1
Line 2 Situation Adv Std Bas
Do not touch the patient!
Analyzing heart rhythm.
Please wait.
DO NOT TOUCH PATIENT
ANALYZING RHYTHM
Repeats until analysis of the patient’s cardiac rhythm
is completed. This prompt will be interrupted if the
AED is ready to shock.
X X
Do not touch the patient
Analyzing Rhythm.
DO NOT TOUCH PATIENT
ANALYZING RHYTHM
Repeats until analysis of the patient’s cardiac rhythm
is completed. This prompt will be interrupted if the
AED is ready to shock.
X
Shock advised. Do not touch
the patient.
SHOCK ADVISED
DO NOT TOUCH PATIENT
Noties the rescuer that a shockable rhythm has been
detected and is preparing to deliver a debrillation
shock (charging).
X X X
Shock not advised SHOCK NOT ADVISED Noties the rescuer when the AED detects a
non-shockable rhythm.
X X X
Analysis interrupted. Stop
patient motion.
ANALYSIS INTERRUPTED
STOP PATIENT MOTION
If the AED detects ECG noise artifacts, stop moving
or touching the patient. Remove other electronic
devices from the vicinity.
X X X
Table A-5: Delivering therapy - G5 semi-automatic
Text display Prompt level
Voice prompt
Line 1
Line 2 Situation Adv Std Bas
Press red ashing button to
deliver shock.
PRESS BUTTON
TO DELIVER SHOCK
Prompts after the AED is fully charged and ready
to deliver the debrillation shock. The red SHOCK
button ashes and the phrase repeats for 30 seconds
or until the SHOCK button is pressed.
X X X
Shock delivered. SHOCK DELIVERED Prompts when the shock is delivered. X X X
Rhythm Changed. Shock
Cancelled.
RHYTHM CHANGED
SHOCK CANCELLED
Noties the rescuer when the AED detects a rhythm
change and cancels the shock.
X X X
A-3Powerheart® G5 Automated External Debrillator 70-01704-01 B