Product Manual
This section presents information about AED product care and
cleaning.
Cardiac Science Corporation provides customer service and
technical support.
◆ To order additional product or accessories, contact Customer Care.
◆ For assistance with product installation or operation, contact
Technical Support. Cardiac Science provides 24-hour support
by telephone. You can also contact Technical Support through
fax or email.
Customer Care
(800) 426 0337 (USA)
(262) 953-3500 (USA and Canada)
care@cardiacscience.com
Technical Support
(800) 426 0337 (USA)
(262) 953-3500 (USA and Canada)
Fax: (262) 798-5236 (USA and Canada)
techsupport@cardiacscience.com
www.cardiacscience.com
Outside the United States and Canada, contact your local
representative.
6 Product Care
Contents
◆ Periodic maintenance ..................................................................... 6-1
◆ Cleaning and care ............................................................................. 6-2
◆ Authorized service ........................................................................... 6-2
Periodic maintenance
Periodically, perform the following tests.
Check the color of the Rescue Ready® indicator.
If the color is…
Do this…
Green No action needed. The AED is ready for a rescue.
Red Refer to Troubleshooting of indicators on page 5-2.
Check that the battery has adequate charge to perform a
rescue:
1. Open the AED lid.
2. If the battery indicator is red, replace the battery.
3. Close the lid.
Check that the voice prompts work and the display is
readable:
1. Open the AED lid.
2. Listen for the voice prompts.
3. Additionally, the display shows text prompts that correspond to
the audio.
4. Close the lid. The voice prompts should stop.
5. Verify that the Rescue Ready indicator returns to green.
If no prompts are heard or they continue after the lid is closed, the
display is not readable, or the Rescue Ready indicator remains red,
there may be an issue with the AED. Contact Cardiac Science Tech-
nical Support, or outside the U.S., your local representative.
Check that the debrillation pads are ready for use and that
the service beep sounds:
1. Open the AED lid.
2. Disconnect the pads connector and remove the pads package.
3. Close the lid.
6-1Powerheart® G5 Automated External Debrillator 70-01704-01 B