Product Manual

Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec.
Not applicable
Power frequency
(50/60 Hz) magnetic
eld
IEC 61000-4-8
3 A/m 80 A/m Power frequency magnetic elds should be at levels no
higher than those characteristic of a typical location in
typical heavy industrial and power plants and the control
rooms of H.V. substations.
Note: U
T
is the a.c. mains voltage prior to application of the test level.
Conducted RF 3 Vrms Not
Applicable
IEC 61000-4-6 150 kHz to 80 MHz outside ISM
bands
a
10 Vrms
150 kHz to 80 MHz in ISM
bands
a
Not
Applicable
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.5 GHz
10 V/m Portable and mobile RF communications equipment
should be used no closer to any part of the AED, including
cables, than the recommended separation distance
calculated from the equation applicable to the frequency
of the transmitter.
Recommended separation distance
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the trans-
mitter in watts (W) according to the transmitter manufac-
turer and d is the recommended separation distance in
meters (m)
b
.
Field strengths from xed RF transmitters, as determined
by an electromagnetic site survey,
c
should be less than the
compliance level in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reection from
structures, objects and people.
a
The ISM (industrial, scientic and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz;
and 40.66 to 40.70 MHz.
b
The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz are intended to decrease the
likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. For this reason, an additional
factor of 10/3 is used in calculating the recommended separation distance for transmitters in these frequency ranges.
c
Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast
and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to xed RF transmitters, an electromagnetic site
survey should be considered. If the measured eld strength in the location in which the AED is used exceeds the applicable RF compliance level above, the AED should
be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the AED.
Electromagnetic Emissions Standards Compliance
Immunity test IEC 60601 test level
Compliance
level
Electromagnetic
enviroment —guidance
D-2 Powerheart® G5 Automated External Debrillator70-01704-01 B