Product Manual
ECG Analysis Algorithm and Rescue Waveform
Waveform and energy levels for preinstalled
debrillation pads (Model number XELAED001)
Table C-1: Ultra-low variable energy waveform
(all values are typical)
Phase 1 Phase 2
Patient’s
Impedance
(Ohms)
Voltage
(Volts)
(V1)
Duration
(ms)
(T1)
Voltage
(Volts)
(V2)
Duration
(ms)
(T2)
Energy
(Joules)
25 1412 3.25 743 3.2 146-197
50 1426 4.50 907 3.2 128-172
75 1431 5.75 968 3.2 116-156
100 1433 7.00 1000 3.2 108-144
125 1435 8.25 1019 3.2 102-136
150 1436 9.50 1031 3.2 97-130
175 1437 10.75 1038 3.2 94-126
Table C-2: Low variable energy waveform
(all values are typical)
Phase 1 Phase 2
Patient’s
Impedance
(Ohms)
Voltage
(Volts)
(V1)
Duration
(ms)
(T1)
Voltage
(Volts)
(V2)
Duration
(ms)
(T2)
Energy
(Joules)
25 1631 3.25 858 3.2 195-263
50 1647 4.50 1047 3.2 170-230
75 1653 5.75 1118 3.2 154-208
100 1655 7.00 1155 3.2 143-193
125 1657 8.25 1176 3.2 135-182
150 1658 9.50 1190 3.2 129-174
175 1659 10.75 1199 3.2 125-168
Table C-3: High variable energy waveform
(all values are typical)
Phase 1 Phase 2
Patient’s
Impedance
(Ohms)
Voltage
(Volts)
(V1)
Duration
(ms)
(T1)
Voltage
(Volts)
(V2)
Duration
(ms)
(T2)
Energy
(Joules)
25 1895 3.25 997 3.2 263-355
50 1914 4.50 1216 3.2 230-310
75 1920 5.75 1299 3.2 208-280
100 1923 7.00 1342 3.2 193-260
125 1925 8.25 1367 3.2 183-246
150 1926 9.50 1383 3.2 174-235
175 1927 10.75 1393 3.2 168-226
Waveform and energy levels for pediatric
debrillation pads
Table C-4: Ultra-low variable energy waveform
(all values are typical)
Phase 1 Phase 2
Patient’s
Impedance
(Ohms)
Voltage
(Volts)
(V1)
Duration
(ms)
(T1)
Voltage
(Volts)
(V2)
Duration
(ms)
(T2)
Energy
(Joules)
25 682 3.25 359 3.2 35-46
50 689 4.50 438 3.2 30-40
75 691 5.75 468 3.2 27-36
100 692 7.00 483 3.2 25-33
125 693 8.25 493 3.2 24-31
150 694 9.50 498 3.2 23-30
175 694 10.75 802 3.2 22-29
Table C-5: Low variable energy waveform
(all values are typical)
Phase 1 Phase 2
Patient’s
Impedance
(Ohms)
Voltage
(Volts)
(V1)
Duration
(ms)
(T1)
Voltage
(Volts)
(V2)
Duration
(ms)
(T2)
Energy
(Joules)
25 791 3.25 416 3.2 46-61
50 798 4.50 508 3.2 40-54
75 801 5.75 542 3.2 37-48
100 802 7.00 560 3.2 34-45
125 803 8.25 570 3.2 32-42
150 804 9.50 577 3.2 31-40
175 804 10.75 581 3.2 30-39
Table C-6: High variable energy waveform
(all values are typical)
Phase 1 Phase 2
Patient’s
Impedance
(Ohms)
Voltage
(Volts)
(V1)
Duration
(ms)
(T1)
Voltage
(Volts)
(V2)
Duration
(ms)
(T2)
Energy
(Joules)
25 915 3.25 481 3.2 62-82
50 924 4.50 588 3.2 54-72
75 927 5.75 628 3.2 49-65
100 929 7.00 648 3.2 46-60
125 930 8.25 660 3.2 43-57
150 931 9.50 668 3.2 41-54
175 931 10.75 673 3.2 40-52
C-2 Powerheart® G5 Automated External Debrillator70-01704-01 B