Product Manual

Table 4: Applicable standards (continued)
Type Detail
Shipping and
transportation
ISTA Procedure 2A
Sensitivity and
specicity of
Rhythm Detection
Shockable Rhythm—VF: Meets IEC 60601-2-4
requirement and AHA recommendation of
Sensitivity of >90%
Shockable Rhythm—VT: Meets IEC 60601-2-4
requirement and AHA recommendation of
Sensitivity of >75%
Non-shockable Rhythm—NSR: Meets IEC
60601-2-4 requirement (>95%) and AHA rec-
ommendation (>99%) of Specicity
Non-shockable—Asystole: Meets IEC 60601-
2-4 requirement and AHA recommendation of
Specicity of >95%
Non-shockable—all other rhythms: Meets IEC
60601-2-4 requirement and AHA recommenda-
tion of Specicity – all other rhythms of >95%
Debrillation pads
Table 5: Preinstalled debrillation pads (model XELAED001)
Parameter Detail
Type Pre-gelled, self-adhesive, disposable, non-polar-
ized (identical pads, which can be placed in either
position) debrillation pads
Applicable age
and weight of
patient
Older than 8 years or heavier than 25 kg (55 lb)
Shelf life 24 months
Disposal Check local regulations for disposal information
Table 6: Preinstalled debrillation pads with CPR feedback
device (model XELAED002)
Parameter Detail
Type Pre-gelled, self-adhesive,
disposable, non-polarized (identical pads, which
can be placed in either position) debrillation
pads with CPR feedback device
Applicable age
and weight of
patient
Older than 8 years or heavier than 25 kg (55 lb)
Shelf life 24 months
Disposal Check local regulations for disposal information
Technical Data
Table 7: Pediatric debrillation pads (model XELAED003)
Parameter Detail
Type Pre-gelled, self-adhesive, disposable, non-polar-
ized (identical pads, which can be placed in either
position) debrillation pads
Applicable age
and weight of
patient
Eight years or younger or 25 kg (55 lb) or lighter
Shelf life 24 months
Disposal Check local regulations for disposal information.
Intellisense
®
battery
(model XBTAED001)
Table 8: Intellisense battery
Parameter Detail
Type Intellisense lithium battery, non-rechargeable
Output voltage 12 VDC (nominal)
Lithium content 9.2 g (approximate)
Disposal Check local regulations for disposal information
Estimated shelf
life*
5 years from date of manufacture
Temperature ranges:
Short term (3 days at either temperature
extreme): -30°C to 65°C
Long term (5 years at either temperature
extreme): 20°C to 30°C
Estimated
operating life**
(new and fully
charged battery)
Shocks: 420 (typical), 250 (minimum), or 16
hours of operating time at 20-30°C, or
Standby: 4 years
* Shelf life is the length of time a battery can be stored prior to installation into an
AED without signicantly aecting its operating life.
** The battery operating life depends on the type of battery, device settings, actual
usage, and environmental factors. The number of shocks is estimated at a 300 VE
energy level with a “three shock stack” followed by 60 seconds of CPR using Basic
prompt settings between each set of shocks.
B-3Powerheart® G5 Automated External Debrillator 70-01704-01 B