Product Manual
Powerheart® AED G3 Plus 9390A and 9390E 70-00914-01 F
6-9
* All values are typical.
**Allowable energy range.
Table 6-4: High Variable Energy Powerheart G3 Waveform (all values are
typical)
Phase 1 Phase 2
Patient’s
Impedance
(Ohms)
Voltage
* (Volts)
Duration*
(MS)
Voltage
* (Volts)
Duration*
(MS)
Energy**
(Joules)
25 1869 3.3 997 3.2 260-351
50 1906 4.5 1220 3.2 230-311
75 1918 5.8 1306 3.2 209-283
100 1925 7.0 1351 3.2 195-263
125 1928 8.3 1378 3.2 184-248
150 1931 9.5 1396 3.2 176-238
175 1933 10.8 1408 3.2 170-230
STAR biphasic waveform