Product Manual
70-00914-01 F Powerheart® AED G3 Plus 9390A and 9390E
6-8
Technical Data
Table 6-2: Ultra-low Variable Energy (150 VE) Powerheart G3 Waveform
Phase 1 Phase 2
Patient’s
Impedance
(Ohms)
Voltage
* (Volts)
Duration*
(MS)
Voltage
* (Volts)
Duration*
(MS)
Energy**
(Joules)
25 1393 3.3 743 3.2 145-196
50 1420 4.5 909 3.2 128-173
75 1430 5.8 973 3.2 116-156
100 1434 7.0 1007 3.2 108-146
125 1437 8.3 1027 3.2 102-138
150 1439 9.5 1040 3.2 98-132
175 1441 10.8 1049 3.2 95-128
Table 6-3: Low Variable Energy (200 VE) Powerheart G3 Waveform
Phase 1 Phase 2
Patient’s
Impedance
(Ohms)
Voltage
* (Volts)
Duration*
(MS)
Voltage
* (Volts)
Duration*
(MS)
Energy**
(Joules)
25 1609 3.3 858 3.2 193-260
50 1640 4.5 1050 3.2 170-230
75 1651 5.8 1124 3.2 155-209
100 1656 7.0 1163 3.2 144-194
125 1660 8.3 1186 3.2 136-184
150 1662 9.5 1201 3.2 130-176
175 1663 10.8 1212 3.2 126-170