Product Manual

70-00914-01 F Powerheart® AED G3 Plus 9390A and 9390E
5-6
Troubleshooting and Maintenance
1. Immediately after opening the AED lid, press and hold the Shock
button and confirm that the Service LED is lit (for the Powerheart
G3 model 9390E only).
2. Release the Shock button (for the Powerheart G3 model 9390E
only).
3. Close the lid.
4. Verify that the STATUS INDICATOR remains RED.
5. Open the lid and confirm that no diagnostic panel indicators are
lit.
6. Close the lid.
7. Verify that the STATUS INDICATOR turns GREEN.
Check the integrity of the case:
Examine the molded case of the AED for any visible signs of stress. If
the case shows signs of stress, contact Cardiac Science Technical
Support (see Contact information on page 1-2) or outside the U.S.,
your local Cardiac Science representative.
!
Caution: Equipment Damage.
When cleaning the device, use one of the following: Isopropyl
Alcohol, Ethanol, a mild soapy water solution, or a 3% hydrogen
peroxide solution.
!
Caution: Equipment Damage.
Keep all cleaning solutions and moisture away from the inside of all
defibrillation pads and cable connector openings.
Authorized repair service
The AED has no user-serviceable internal components. Try to resolve
any maintenance issues with the AED by using the Troubleshooting
Table presented in this chapter. If you are unable to resolve the
problem, contact Cardiac Science Technical Support (see Contact
information on page 1-2) or outside the U.S., your local Cardiac
Science representative.
!
WARNING! Shock Hazard.
Do not disassemble the AED. Failure to observe this warning can
result in personal injury or death. Refer maintenance issues to Cardiac
Science authorized service personnel.
Note: The warranty will be void upon unauthorized disassembly or
service of the AED.