Product Manual
Powerheart® AED G3 Plus 9390A and 9390E 70-00914-01 F
5-1
5 Troubleshooting and
Maintenance
Contents
◆ Self-tests 5-1
◆ Indicator troubleshooting table 5-3
◆ Scheduled maintenance 5-4
◆ Authorized repair service 5-6
◆ Frequently Asked Questions 5-7
This section presents information about the AED diagnostics self-tests,
maintenance, and service indications.
Self-tests
The AED has a comprehensive self-test system that automatically tests
the electronics, battery, pads, and high voltage circuitry. Self-tests are
also activated every time you open and close the AED lid.
When performing the self-tests, the AED completes the following steps
automatically:
1. Turns itself on, and the Status Indicator changes to red.
2. Performs the self-test.
3. If successful, the Status Indicator reverts to green.
4. Turns itself off if the lid is closed.
There are three types of automatic self-tests:
◆ The daily self-test checks the battery, pads, and the electronic
components.