Product Manual
70-00914-01 F Powerheart® AED G3 Plus 9390A and 9390E
3-14
Getting Started
“Rhythm changed.
Shock cancelled.”
RHYTHM CHANGED
SHOCK CANCELLED
When the device is prepared to shock then detects a change in rhythm and therefore
cancels the shock.
“Remove cable to
continue rescue.”
REMOVE CABLE
TO CONTINUE RESCUE
When a serial communication cable is connected to the AED during a rescue, the phrase
repeats until the cable is disconnected.
“Communications
mode”
COMMUNICATIONS MODE When the lid is open and the serial communication cable is plugged into the AED.
“Service required” SERVICE REQUIRED Occurs after the self-tests determine that the AED is not functioning properly. The
prompt “SERVICE REQUIRED” will be heard when the lid is opened. The red SERVICE
indicator will illuminate. After closing the lid, an alarm beep will be heard until the
battery is removed or becomes completely depleted.
Table 3-8: Other prompts (continued)
Voice prompt Text display Situation