Product Manual

Powerheart® AED G3 Plus 9390A and 9390E 70-00914-01 F
3-13
RescueCoach™ voice prompts and text display
“Press pads firmly to patient’s
bare skin.
PRESS PADS FIRMLY
TO BARE SKIN
Prompts when better pad connectivity to the patient’s skin is required because
impedance is too high.
Table 3-8: Other prompts
Voice prompt Text display Situation
“Battery low BATTERY LOW Occurs once when the battery voltage becomes low, although a rescue can continue for
approximately 9 more shocks. When the battery is too low to do a rescue, the following
will occur:
BATTERY LOW shows on the LCD
Smartgauge battery status indicator turns red
AED beeps once every 30 seconds while the lid is closed
You must replace the battery before continuing with the rescue. If completely depleted,
all AED activity will terminate.
Analysis interrupted.
Stop patient motion.
ANALYSIS INTERRUPTED
STOP PATIENT MOTION
When the AED detects ECG noise artifact, stop moving or touching the patient. Remove
other electronic devices within a 5 meter radius.
“Open lid to continue
rescue.
OPEN LID
TO CONTINUE RESCUE
When the lid is inadvertently closed during a rescue, this prompt will repeat for 15
seconds.
Table 3-7: Pad issues
Voice prompt Text display Situation