Product Manual

70-00914-01 F Powerheart® AED G3 Plus 9390A and 9390E
3-10
Getting Started
Table 3-5: Delivering shock - Fully automatic
Voice prompt Text display Situation
“Shock will be delivered in SHOCK IN: After the AED is fully charged and ready to deliver the defibrillation shock. The
SHOCK will automatically be administered approximately three seconds after the
end of the voice prompt.
Three THREE Prompts approximately three seconds prior to delivering shock.
Two TWO Prompts approximately two seconds prior to delivering shock.
“One ONE Prompts approximately one second prior to delivering shock.
“Shock delivered” SHOCK DELIVERED Prompts when the shock is delivered.
Table 3-6: CPR prompts
Voice prompt Text display Situation
Note: The AED is shipped from the factory with ENHANCED MODE defaulted ON. The Medical Director may modify the CPR options in
MDLink. ENHANCED CPR prompts are listed in this table. Except where noted, prompts apply both to compressions-only CPR and
traditional CPR (compressions and breaths).
“It is now safe to touch the patient. NOW SAFE
TO TOUCH THE PATIENT
Advises the rescuer that it is safe to touch the patient:
After the AED delivers a shock
After the AED detects a non-shockable cardiac rhythm