Product Manual
70-00914-01 F Powerheart® AED G3 Plus 9390A and 9390E
3-8
Getting Started
“When patient’s chest and torso are exposed,
remove square foil package from lid of AED.”
WHEN CHEST IS BARE
REMOVE FOIL PACKAGE
Prompts the rescuer to remove the pads from AED lid.
“Tear open foil package across dotted line and
remove pads.”
TEAR OPEN PACKAGE
REMOVE PADS
Prompts the rescuer to open the pad package and remove pads.
“Next, separate one of the white pads
completely from blue plastic liner. Begin
peeling from the tabbed corner.”
PEEL ONE PAD FROM
BLUE PLASTIC LINER
Repeats every 3 seconds until the pads are separated. If a pad has
been peeled before the prompt starts, this prompt will be skipped.
This prompt will be interrupted when pad is peeled.
“Firmly place the pad without the liner on the
patient, exactly as illustrated. This pad can be
placed on either of the two locations shown.”
PRESS PAD FIRMLY
TO CHEST AS SHOWN
Prompts the rescuer to place one pad on the patient.
“Next, peel the blue plastic liner off of the
second white pad.”
PEEL SECOND PAD OFF
BLUE PLASTIC LINER
Prompts the rescuer to remove the liner from the second pad.
“Firmly place the second pad on the opposite
location, exactly as illustrated.”
PRESS PAD FIRMLY
AS SHOWN
Repeats until second pad placement is sensed. If the pad is placed
before prompt starts then this prompt will be skipped. This prompt
will be interrupted when second pad is placed.
Table 3-2: Preparation (continued)
Voice prompt Text display Situation