Product Manual
70-00914-01 F Powerheart® AED G3 Plus 9390A and 9390E
3-6
Getting Started
Setting the AED internal clock
For US models, the internal clock is preset to Central Standard Time.
You can reset it to your local date and time. To set the clock, you need a
Windows XP or newer PC, Rescuelink software installed, and the AED
serial cable connected to the PC.
To set the clock:
1. Ensure that the PC is set at the correct local time and date.
2. Open the lid of the AED and run the Rescuelink software on the
PC.
3. Connect the cable to the serial port on the AED.
4. Verify that the voice prompt states “Communications Mode”.
5. Click Communications on the main menu. Select AED Date and
Time.
6. Click on the Get button to review the current time in the AED.
7. If the time and date is incorrect, click Set to set new time and date.
The AED date and time will automatically be updated to the PC’s
time and date.