Product Manual
70-00914-01 F Powerheart® AED G3 Plus 9390A and 9390E
3-4
Getting Started
longer than 60 seconds, the first rescue will be terminated and a second
rescue will begin upon insertion of battery.
Note: When the battery is depleted, neither the LED nor the text
display illuminates.
Pads indicator
The Pads LED lights up when the pads are:
◆ Not properly connected to the AED
◆ Not within operational specifications
(cold, dried, damaged)
◆ Disconnected from the patient during a
rescue
Service indicator
The Service LED lights up when the AED detects
an error that cannot be corrected by the self test.
Contact Cardiac Science Technical Support (see
Contact information on page 1-2) or outside the
U.S., your local Cardiac Science representative.
Shock button
For the Powerheart G3 model 9390E only: The
AED has one button called the Shock button. The
word Shock and the shock button LED will
illuminate red when the AED is ready to deliver a
defibrillation shock to the patient.