Product Manual

Powerheart® AED G3 Plus 9390A and 9390E 70-00914-01 F
2-5
Detection Rate and is selectable between 160 and 300 bpm or, “NO
THERAPY FOR SVT” can be selected via MDLink Software by the
Medical Director.
Rescue protocol
The AED rescue protocol is consistent with the guidelines
recommended by the AHA/ERC 2010 Guidelines for Resuscitation
and Emergency Cardiac Care.
Upon detecting a shockable cardiac rhythm, the AED advises the
operator to press the SHOCK button (9390E only) to deliver a
defibrillation shock followed by directions to perform 2 minutes of
CPR.
For the Powerheart AED G3 Automatic, upon detecting a shockable
rhythm, the AED will automatically deliver a defibrillation shock
followed by directions to perform 2 minutes of CPR.
STAR
®
biphasic waveform
The STAR
®
Biphasic Waveform is designed to measure the patient’s
impedance and deliver a customized shock. This allows the delivery of
an optimized energy level to each patient. The energy levels for the
Powerheart G3 AED are available in three different defibrillation shock
levels.
The Ultra-Low Energy (150 VE), Low Energy (200 VE), and High
Energy (300 VE) shocks are variable energy. The actual energy is
determined by the patients impedance. See Table 2-1 on page 2-6,
Table 6-2 on page 6-8, Table 6-3 on page 6-8, and Table 6-4 on
page 6-9 for additional information.
Rescue protocol