Product Manual
Powerheart® AED G3 Plus 9390A and 9390E 70-00914-01 F
1-19
Electromagnetic emissions standards compliance
Recommended separation distances between portable and mobile RF communications
equipment and the AED
The AED is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the AED can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the AED as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
outside ISM bands
d = 1.2
150 kHz to 80 MHz in
ISM bands
d = 1.2
80 MHz to 800 MHz
d = 1.2
800 MHz to 2.5 GHz
d = 2.3
0.01 0.12 0.12 0.12 0.23
0.1 0.38 0.38 0.38 0.73
1 1.2 1.2 1.2 2.3
10 3.8 3.8 3.8 7.3
100 12 12 12 23
P
P
P
P