Product Manual
70-00914-01 F Powerheart® AED G3 Plus 9390A and 9390E
1-14
Electromagnetic emissions standards compliance
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
61000-4-11
<5% U
T
(>95% dip in U
T
) for 0.5
cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec.
Not applicable
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
3 A/m 80 A/m Power frequency magnetic fields should
be at levels no higher than those
characteristic of a typical location in
typical heavy industrial and power
plants and the control rooms of H.V. sub-
stations.
Note:
U
T
is the a.c. mains voltage prior to application of the test level.
Immunity test IEC 60601 test level Compliance level
Electromagnetic environment—
guidance