Product Manual

Powerheart® AED G3 Plus 9390A and 9390E 70-00914-01 F
1-13
Electromagnetic emissions standards compliance
Guidance and manufacturers declaration—electromagnetic immunity
The AED is intended for use in the electromagnetic environment specified below. The customer or the user of the AED should
assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic environment—
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
Not applicable
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
Not applicable