Product Manual
70-00914-01 F Powerheart® AED G3 Plus 9390A and 9390E
1-4
Product Information and Safety
!
Caution
This alert identifies hazards that may cause minor personal injury,
product damage, or property damage.
Safety alert descriptions
The following is a list of Powerheart G3 AED safety alerts that appear
in this section and throughout this manual.
Read and understand these safety alerts before operating the AED.
!
Caution: Read this Operator and Service Manual carefully.
It contains information about your safety and the safety of others.
Become familiar with the controls and how to use the AED properly
before operating the product.
!
DANGER! Fire and Explosion Hazard
To avoid possible fire or explosion hazard, do not operate the AED:
• In the presence of flammable gases
• In the presence of concentrated oxygen
• In a hyperbaric chamber
!
WARNING! Shock Hazard and Possible Equipment Damage
Defibrillation shock current flowing through unwanted pathways
is potentially a serious electrical shock hazard and potential
damage to the equipment. To avoid this hazard during
defibrillation abide by all of the following:
• Do not use in standing water or rain. Move patient to dry
area
• Do not touch the patient, unless performance of CPR is
indicated
• Do not touch metal objects in contact with the patient
• Keep defibrillation pads clear of other pads or metal parts
in contact with patient
• Disconnect all non-defibrillator proof equipment from
the patient before defibrillation
!
WARNING! Battery is Not Rechargeable.
Do not attempt to recharge the battery. Any attempt to recharge the
battery may result in an explosion or fire hazard.