Specifications

23
High Performance Liquid Chromatograph
Column Oven / Fraction Collector
The CTO-10AS
VP
is a space-saving, compact and affordable oven with a block heating mechanism.
The CTO-10AS
VP
controls temperature from −15°C below room temperature to 60°C, and
accommodates two manual injectors.
CTO-10ASVP
Specifications
Block heating
Electronic cooling
4 to 80°C
±0.1°C
−15 to 60°C, room temperature
25 cm (2 columns max.)
Change of temperature setting
Leak sensor, temperature fuse, temperature upper limit
W130 × D420 × H415 mm, 12 kg
AC 110 V, 230 V, 120 VA, 50/60 Hz
Type
Cooling method
Temperature setting range
Temperature control precision
Temperature control range
Applicable columns
Function
Safety features
Dimensions, Weight
Power requirements
CTO-10AS
VP (228-45059-XX)
FRC-10A
A Fraction Collector for a Wide Variety of Fractionation Modes
The FRC-10A can be used over a wide range of flow rates, covering small and large-scale
preparative work. It flexibly adapts to various applications, such as simple, manual collection
performed while viewing chromatograms, and advanced, continuous and automated
preparative separation and collection performed in combination with an autosampler and
Convenient Fraction Simulation
Fraction simulation can be performed using LabSolutions, so the optimization of fractionation conditions is very simple.
Minimal Influence of Variations in Elution Time
Even if the elution time changes due to the influence of fluctuations in room temperature or the composition of the mobile phase, it is still
possible to accurately perform fractionation by catching the target component with special parameters. This function is indispensable for
continuous automatic preparative separation.
Specifications
FRC-10A (228-45070-xx)
Arm-movement X–Y system
16 to 144 (depending on the type of rack used)
Solenoid valve (fraction-collector head with valve) or direct through nozzle (fraction-collector head)
150 mL/min
Basic mode (using initial parameters), and Time-Program mode (14 different functions available)
Possible with Sample Cooler L (228-45064-xx)
4 to 35°C
W260 × D420 × H280 mm, 15 kg
AC 110 V, 230 V, 100 VA, 50/60 Hz
Drive system
Maximum number of fractions
Collection method
Maximum flow rate
Fraction modes
Cooling function
Ambient temperature range
Dimensions, weight
Power requirements