Specifications

22
Solvent Delivery Unit / Autosamplers
Specifications
LC-10Ai (228-45089-xx)
Serial dual plunger
Primary side: 47 µL; Secondary side: 23 µL
27.4 MPa
0.001 to 9.999 mL/ min
Within ±2% or ±2 µL /min., whichever is larger (0.1 to 5.0 mL/ min.)
0.3% max. (RSD: 0.1% max.)
Possible
Syringe or rinsing pump (228-39625-91) used
4 to 35°C
W260 × D420 × H140 mm, 10 kg
AC 110 V, 230 V, 100 VA, 50/60 Hz
Solvent-delivery method
Plunger capacity
Maximum discharge pressure
Flow rate setting range
Flow rate accuracy
Flow rate precision
Constant-pressure delivery
Plunger rinse mechanism
Operating temperature range
Dimensions, weight
Power requirements
* Use the preparative mixer (228-20600-91) if the flow rate is greater than or equal to 10 mL/min.
SIL-10AF / 10AP / 10Ai
This bio-inert solvent delivery unit incorporates a serial dual plunger and offers low-pulsation
performance from an optimized cylinder volume. It can be used together with the SIL-10Ai and
SPD-20A (with bio-inert cell) to construct a high-performance bio-inert LC system. The LC-10Ai
uses PEEK resin in liquid contact parts and is ideal for the analysis of physiologically active
substances and metal ions. Resistance to acids, bases, and high-concentration aqueous NaCl
solutions is even higher than with stainless steel.
LC-10Ai
Specifications
* SIL-10AF / 10AP / 10Ai can not control from CBM-20Alite.
LC-10Ai Solvent Delivery Unit for Bio-inert HPLC System
The SIL-10AF and SIL-10AP injectors use the fixed-loop injection method. They can also perform sample
pretreatment, including dilution and mixing, at a high speed. The SIL-10AP is a preparative autosampler
that can inject up to 5 mL while offering the same level of performance and functionality as the SIL-10AF.
This bio-inert autosampler uses PEEK resin in parts that contact liquids. It can be used for the
analysis of physiologically active substances and metal ions.
SIL-10AF / 10AP
SIL-10Ai
Versatile Autosamplers
Bio-inert Autosampler
1 to 50 µL (standard) 1 to 400 µL (option)
1 to 2,000 µL (option) 1 to 5,000 µL (option)
100 with 1.5 mL vials (60 with optional cooler) 80 with 4 mL vials (50 with optional cooler)
25 with 13 mL vials (not applicable to SIL-10Ai) 192 with two 96-well microtiter plates
Optional Sample Cooler S (228-45063-xx) or L (228-45064-xx) Block-heating /cooling method 4 to 70°C
Not specified
Not specified
30 max. per sample
Set freely before and after sample injection
4 to 35°C
1 to 5,000 µL (standard)
1 to 400 µL (option) 1 to 2,000 µL (option)
1 to 50 µL (standard)
1 to 250 µL (option)
Injection method
Injection-volume setting range
Number of samples processed
Injection-volume accuracy
Injection-volume precision
Sample carryover
Number of repeated injections
Needle rinsing
Sample cooler
Operating pH range
Operating temperature range
Dimensions, weight
Power requirements
RSD: 0.5% max. (10 µL injection, standard mode) RSD: 0.5% max. (10 µL injection, standard mode)RSD: 1% max. (10 µL injection)
SIL-10AF (228-45056-xx) SIL-10AP (228-45057-xx) SIL-10Ai (228-45075-xx)
Loop injection, variable injection volume
Main unit: W260 × D420 × H280 mm, 19 kg Syringe unit: W100 × D150 × H280 mm, 4 kg
pH1 to pH10
AC 110 V, 230 V, 100 VA, 50/ 60 Hz