Technical information

Users Manual: DFM19HBV
2007 Richardson Electronics GmbH
Division A.C.T. Kern
8
WARNING:
Risk of Suffocation — Dispose of the packaging material, observing the applicable waste-
control regulations and keeping it out of children's reach.
CAUTION:
Personal Injury — All devices used must comply with the applicable IEC/EN or ISO standard
(PCs, for instance, must comply with IEC 60950). This applies to custom designs in particular
CAUTION:
Personal Injury, Equipment Damage — Check that the ambient conditions specified by the
manufacturers of the different devices are ensured.
Before cleaning the flat-screen display, turn it off and remove the 24VDC connector or the
power supply connection cable. Do not spray liquid cleaning agents directly onto the display.
Spray the cleaning agent onto a cloth and wipe the display clean without exerting any
pressure.
Do not place any objects on the signal cables. Route the signal cables and the power cable
to the flat-screen display such that they do not present a stumbling hazard.
CAUTION:
Equipment Damage — Do not set up the flat-screen display in the immediate vicinity of a
window. Rain, water, humidity and sunlight may damage the flat-screen display.
Choose a flat, solid surface to set up the flat-screen display.
Do not exert pressure on the LCD. Excessive pressure may permanently damage the display.
For safe storage of the flat-screen display, a temperature between -20°C and +60°C must be
ensured. Temperatures above or below this range may permanently damage the flat-screen
display. Always store your flat-screen display in its original shipping box.
WARNING:
Personal Injury — During intracardiac application, it is recommended to connect the
individual devices to the room’s potential equalization system via separate potential
equalization cables.
For intracardiac application, the application guidelines for medical electrical devices set
forth in the German standard VDE 0753, part 2, must be observed.