LV-7575 Expand Serial Command Functional Specifications

LV-7575 Expand Serial Command Functional Specifications
58/71
Unacceptable “Error Code” [CR]
9.12.2 CR_PRESSURE Command
Command “CR_PRESSURE” [CR]
Details
Get value from Air Pressure sensor.
Fllowing is the formula to work out Air Pressure from the given value (“Vn”
represents the value):
Air Pressure (hPa) =5*Vn / 1024 – 0.204/ 0.00459 + 150
Obtains accuracy of +/- 2%
Acceptable “000_%1” [CR]
%1 “0000” – “1023”
Response
Unacceptable “Error Code” [CR]
9.12.3 CR_SIGNAL Command
Command “CR_SIGNAL” [CR]
Details Get status whether there is any signal or not.
Acceptable “000_%1” [CR]
%1
“ON” -------------------------- There is signal
“OFF” ------------------------- There is no signal
Response
Unacceptable “?” [CR]
9.12.4 CR_VMUTE Command
Command “CR_VMUTE” [CR]
Details Get setting status of No Show.
Acceptable “000_%1” [CR]
%1
“ON” -------------------------- No show is enabled
“OFF” ------------------------ No show is disabled
Response
Unacceptable “?” [CR]
9.12.5 CR_FREEZE Command
Command “CR_FREEZE” [CR]
Details Get setting status of Freeze.
Acceptable “000_%1” [CR]
%1
“ON” ---------------------------- Freeze is enabled
“OFF” -------------------------- Freeze is disabled
Response
Unacceptable “?” [CR]