TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
NOTEQual triggers the oscilloscope when the runt pulse width is true for a time
period greater than or less than (not equal to) the time period specified in
TRIGger:A:PULse:RUNT:WIDth, within a ±5% tolerance.
TRIGGER:A:PULSE:RUNT:WHEN MORETHAN
sets the runt trigger to occur when the oscilloscope detects a runt pulse wider
than the specified width.
TRIGger:A:PULse:RUNT:WIDth
Sets or returns the runt pulse width time period value. This command is
equivalent to entering a value in the runt pulse trigger menu’s Trigger When
side menu.
Trigger
TRIGger:A:PULse:RUNT:WIDth <NR3>
TRIGger:A:PULse:RUNT:WIDth?
<NR3> is a floating point value with exponent that sets the runt pulse trigger time
value. This argument has a range of 39.6E--9 (39.6 ns) to 10.0E0 (10 s), in
increments of 13.2 ns. Values that are not an increment of 13.2 ns are rounded to
the nearest correct value.
TRIGGER:A:PULSE:RUNT:WIDTH 15E-6
sets the minimum width of the pulse runt trigger to 15.0084 s.
TRIGger:A:PULse:SLEWRate? (Query Only)
Returns the slew rate trigger settings.
Trigger
TRIGger:A:PULse:SLEWRate?
TRIGGER:A:PULSE:SLEWRATE?
might return TRIGGER:A:PULSE:SLEWRATE:DELTATIME 0.0000011880;
THRESHOLD:HIGH -0.800000000;LOW -2.000000000;
:TRIGGER:A:PULSE:SLEWRATE:WHEN FASTERTHAN
Examples
Group
Syntax
Arguments
Examples
Group
Syntax
Examples