TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 281
TRIGger:A:PULse:RUNT:THReshold:LOW
Sets or returns the runt pulse signal low threshold voltage value. This command
is equivalent to setting the threshold in the runt pulse Thresholds side menu.
Trigger
TRIGger:A:PULse:RUNT:THReshold:LOW <NR3>
TRIGger:A:PULse:RUNT:THReshold:LOW?
<NR3> is a floating point value with exponent that sets the runt pulse low
threshold voltage value.
TRIGGER:A:PULSE:RUNT:THRESHOLD:LOW -700.0E-3
sets the runt pulse low threshold voltage value to --700 mV.
TRIGger:A:PULse:RUNT:WHEn
Sets or returns the runt pulse width condition on which to trigger the oscillo-
scope.
Trigger
TRIGger:A:PULse:RUNT:WHEN { OCCurs | LESSThan | MOREThan |
EQual | NOTEQual }
TRIGger:A:PULse:RUNT:WHEN?
OCCurs triggers the oscilloscope on any runt signal regardless of width.
LESSThan triggers the oscilloscope when the runt pulse width is true for a time
period less than the time period specified in TRIGger:A:PULse:RUNT:WIDth.
MOREThan triggers the oscilloscope when the runt pulse width is true for a time
period more (greater) than the time period specified in TRIGger:A:PULse:
RUNT:WIDth.
EQual triggers the oscilloscope when the runt pulse width is true for a time
period equal to the time period specified in TRIGger:A:PULse:RUNT:WIDth,
within a ±5% tolerance.
Group
Syntax
Arguments
Examples
Group
Syntax
Arguments