Specifications
Command Descriptions
2- 278
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
TRIGger:A:PULse? (Query Only)
Returns the pulse trigger parameters.
Trigger
TRIGger:A:PULse?
TRIGGER:A:PULSE?
might return TRIGGER:A:PULSE:CLASS WIDTH;SOURCE CH3;POLARITY
EITHER;WIDTH:WHEN MORETHAN;WIDTH 0.0000011880;:TRIG-
GER:A:PULSE:RUNT:THRESHOLD:HIGH 0.8000000000;LOW
2.0000000000;:TRIGGER:A:PULSE:RUNT:WHEN LESSTHAN;WIDTH
0.0000011880;:TRIGGER:A:PULSE:SLEWRATE:DELTATIME
0.0000011880;THRESHOLD:HIGH 0.8000000000;LOW 2.0000000000;
:TRIGGER:A:PULSE:SLEWRATE:WHEN FASTERTHAN.
TRIGger:A:PULse:CLAss
Sets or returns the pulse trigger class value. This command is equivalent to
selecting Class in the Trigger menu when the trigger Type is set to Pulse.
Trigger
TRIGger:A:PULse:CLAss { WIDth | RUNT | SLEWRate }
TRIGger:A:PULse:CLAss?
WIDth sets the oscilloscope pulse trigger class to width. The width mode triggers
the oscilloscope when the source signal meets specified pulse width constraints.
RUNT sets the oscilloscope pulse trigger class to runt. The runt mode triggers the
oscilloscope when the source signal meets specified pulse threshold and optional
width constraints.
SLEWRate sets the oscilloscope pulse trigger class to slew rate. The slew rate
mode triggers the oscilloscope when the source signal meets specified slew rate
constraints.
TRIGGER:A:PULSE:CLASS WIDTH
sets the oscilloscope pulse trigger class to width trigger.
Group
Syntax
Examples
Group
Syntax
Arguments
Examples