Specifications
Command Descriptions
2- 222
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
MEASUrement:MEAS<x>:SOURCE2
Sets or returns the source to measure “to” when taking a delay measurement or
phase measurement. This is equivalent to setting the source in the Delay from
Selected Wfm side menu or the Phase from Selected Wfm side menu.
Measurement
MEASUrement:MEAS<x>:SOURCE2 { CH<x> | MATH | REF<x> }
MEASUrement:MEAS<x>:SOURCE2?
CH<x> is an input channel.
MATH is the math waveform.
REF<x> is a reference waveform.
MEASUREMENT:MEAS4:SOURCE2 CH1
sets channel 1 as the delay “to” source when making delay measurements.
MEASUREMENT:MEAS2:SOURCE2?
might return MATH.
MEASUrement:MEAS<x>:STATE
Turns the measurement specified by <x> on or off. The source specified by
MEASUrement:MEAS<x>:SOURCE1 must be active for the measurement to be
displayed. The source can be activated using the SELect:CH<x> command.
Measurement
MEASUrement:MEAS<x>:STATE { OFF | ON | <NR1> }
MEASUrement:MEAS<x>:STATE?
OFF or <NR1> = 0 turns the specified measurement off. You can also turn the state
off by turning the source off.
ON or <NR1> ¸ 0 turns the specified measurement on.
Group
Syntax
Arguments
Examples
Group
Syntax
Arguments