Owner's manual

45
7-2. Principle of measurement
7-2-1. Oscillating U-tube Cell
When a measurement cell filled with liquid or gas is oscillated spontaneously, resultant oscillating cycle
varies with the density of the sample in the measurement cell. The oscillating cycle T is expressed
follows.
T
dVc Mc
K
=
+
2π
where,
d = Density of sample in measurement cell
Vc = Volume of sample in measurement cell, or measurement cell internal volume
Mc = Mass of measurement cell
K = Constant
d
K
Vc
T
Mc
Vc
=
4
2
2
π
Supposing that the oscillating cycles Ta and Tw are obtained through measurement of reference
materials of known density (density da and dw), the facto value F is calculated as follows.
F
K
Vc
d d
T T
a
w
a w
= =
4
2 2 2
π
Accordingly, the density “d” of unknown sample can be calculated according to the following formula by
measuring its oscillating cycle T.
d
d F T T
a a
=
( )
2
2
CANNON
®
D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA