Specifications
Section 6. Maintenance and Troubleshooting 
6-4 
C.  Make sure the datalogger is connected to the modem, and the modem is 
properly configured and cabled (Section 9). 
At the computer: 
D.  Make sure the Station File is configured correctly (LoggerNet or PC400 
Manual). 
E.  Check the cable(s) between the serial port and the modem. If cables have 
not been purchased through Campbell Scientific, check for the following 
configuration using an ohm meter: 
25-pin serial port: 
computer 
end 
modem 
end 
2 2 
3 3 
7 7 
20 20 
9-pin serial port: 
computer 
end 
modem 
end 
2 3 
3 2 
4 20 
5 7 
F.  Make sure the modem is properly configured and cabled (Section 9). 
G.  If still no response, call Campbell Scientific. 
6.2.3 NaN Displayed in a Variable 
A.  Make sure the battery voltage is between 9.6 and 16 VDC. 
B.  Verify the sensor is wired to the analog channel specified in the 
measurement instruction or Short Cut .FSL file (single-ended channels are 
not labeled on the older CR10 silver-colored wiring panels and are 
numbered sequentially starting at 1H; i.e. 1L is single-ended channel 2). 
C.  Make sure the Range parameter in the measurement instruction covers the 
full scale voltage output by the sensor. 
6.2.4 Unreasonable Results Displayed in a Variable 
A.  Inspect the sensor for damage and/or contamination. 
B.  Make sure the sensor is properly wired to the datalogger. 
C.  Check the multiplier and offset parameters in the measurement 
instruction. 