Instruction manual

Model 109SS Temperature Probe
For CR6 datalogger:
Therm109(T109_C,1,U1,U10,20000,_60Hz,1.0,0.0)
For CR800, CR1000, CR3000, and CR5000 dataloggers:
Therm109(T109_C,1,1,1,20000,_60Hz,1.0,0.0)
A setting time parameter is not available for CR200(X) series dataloggers.
9. Troubleshooting and Maintenance
All factory repairs and recalibrations require a returned material
authorization (RMA) and completion of the “Declaration of
Hazardous Material and Decontamination” form. Refer to the
Assistance page at the beginning of this manual for more
information.
9.1 Troubleshooting
Symptom: Temperature is reported as NAN, INF, or incorrect temperature.
Verify wire leads are connected to the terminals specified in the
Therm109() instruction: red to single-ended analog input (SE or U), black
to switched excitation (VX/EX or U), and purple to ground ( ).
Symptom: Incorrect temperature is reported.
Verify the Mult and Offset arguments in
Therm109() are correct for the
desired units (Section 7.2, Datalogger Programming). Check the cable for
signs of damage and possible moisture intrusion.
Symptom: Unstable temperature is reported.
Probably a result of electromagnetic interference. Try using the _50Hz or
_60Hz Integ or Fnotch options, and/or increasing the settling time as
described in Section 8.3, Electrically Noisy Environments, and Section 8.4,
Long Cable Lengths. Ensure the clear wire is connected to datalogger
ground, and the datalogger is properly grounded.
9.2 Maintenance
The 109SS probe requires minimal maintenance. For air temperature
measurements, check the radiation shield monthly to make sure it is clean and
free from debris. Periodically check cabling for signs of damage and possible
moisture intrusion.
9.3 Calibration
Calibration of the 109SS probe is not necessary unless the application requires
removal of the thermistor interchangeability offset described in Section 6,
Specifications.
10