Instruction manual
Model 109SS Temperature Probe
Stainless-Steel Sheath
Diameter: 0.16 cm (0.063 in)
Length: 5.84 cm (2.3 in)
Overmolded Joint
Diameter: 1.02 cm (0.40 in)
Length: 4.24 cm (1.67 in)
Cable: Santoprene
®
, 0.220 in diameter
Cable/Probe Connection: ATUM
TM
heat shrink
Macromelt
®
overmolded joint
Weight: 0.2 lb per 10.5 ft cable
1
See FIGURE 6-1, Worst-case probe and measurement errors, and FIGURE 6-2 Steinhart-Hart
linearization error. Overall probe accuracy is a combination of thermistor interchangeability,
bridge-resistor accuracy, and error of the Steinhart-Hart equation . Interchangeability is the
principle component error and is predominantly offset. Offset can be determined with a single-
point calibration. Offset can be entered in the Therm109() instruction Offset parameter. The
bridge resistor has a 0.1% tolerance with a 10 ppm temperature coefficient. At temperature
extremes, the possible error in the CR200(X) series datalogger measurement may be greater than
the error that exists in the probe.
FIGURE 6-1. Worst-case probe and measurement errors
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
-50 -40 -30 -20 -10 0 10 20 30 40 50 60 70
Possible Error Degrees C
Temperature Degrees C
Worst Case Errors in 109 Temperature Measurement
Thermistor
Tolerance
CR200 Bridge
Measurement
Error (0.06% of
reading + 2.4
mV)
CR200
Resolution (0.6
mV)
5