Specifications

Section 2. Hardware Installation
FIGURE 2.6-2. Grounding to Lightning Rod Clamp
3. Strip 2.54 cm (1”) from both ends of the 9” (23 cm) piece of 10 AWG
green ground wire. Insert one end into the enclosure ground lug located at
the top back of the enclosure. Put the other end in the same clamp
opening as the main grounding wire and tighten down the screw (see
Figure 2.6-2).
4. Insert the lightning rod into the empty clamp opening. The milled flat side
of the lightning rod should face the clamping screw. Tighten the screw to
hold the rod firmly in place.
2-42