Instruction manual
WS-PRO2 Weather Station Manual
GT
27176
Page 60 of 61
reconnecting the battery plug. Keypad should power up and
the message “HELLO” seen on the display.
• Call Campbell Scientific if still no response.
• No Response from Datalogger through SC32A or Modem
Peripheral
At the datalogger:
• Make sure the battery has been installed and the power
switch set to “ON” (section 4.1.1 – 4.1.3).
• Use a voltmeter to measure the voltage on the 12V and G
terminals (see Figure 17); the voltage must be between 9.6
and 16 VDC.
• Use a voltmeter to measure the voltage on the 5V and G
terminals (see Figure 17); the voltage must be between 4.9 –
5.1 VDC.
• Make sure the datalogger is connected to the
communication peripheral, and the communication
peripheral properly installed and configured (section 3.3).
• At the computer:
• Make sure calling software is properly configured (PC200W,
VisualWeather, or LoggerNet).
• Check the cable(s) between the serial port and the modem.
If cables have not been purchased through Campbell
Scientific, check for the following configuration using an ohm
meter:
• 25-pin serial port:
computer end modem end
2 2
3 3
7 7
20 20
9-pin serial port:
computer end modem end
2 3
3 2
4 20
5 7
• Make sure the communication device at the computer is
properly configured and cabled (section 3.3).
• Call Campbell Scientific if still no response.
-99999 Displayed in an Input Location