Instruction manual
HS2 and HS2P 
If the   button is pressed when the GPS does not have a valid GPS sync, a 
warning message is display. 
This message warns that the current measurement may not be stored with an 
accurate date or time.  To continue and store data without GPS information, 
press  . To return to the main screen (p. 14) without storing, press  . 
10.  Troubleshooting 
Problem  Explanation / Recommendation 
Display shows “SENSOR 
TIMEOUT” 
continuously. 
This message indicates that the HydroSense II 
display has not received a response from the 
sensor.  Check that the connectors are mated 
correctly, the pins are clean, and the connector 
has not been damaged.  If this problem persists, 
please contact Campbell Scientific.  
Display shows “NO GPS 
LOCK. STORE 
ANYWAY?” when 
attempting to store data. 
This warning indicates that the GPS has not 
achieved synchronization with the GPS satellite 
network.  This may occur because the GPS has 
been turned off, does not have a clear view of the 
sky or simply has not been turned on long enough 
to achieve synchronization. If selecting OK in 
response to this query, data will be stored without 
positional information and the date and time 
information may be inaccurate.  
The volumetric water 
content is reported as  
“---”. 
“---” indicates that the water content sensor was 
unable to determine the volumetric water content 
for these soil conditions. Try taking the 
measurement a few more times or try another 
location close by.  For extreme or special soils, 
period can sometimes be used in conjunction 
with a soil specific calibration to yield usable 
results.  
The main screen shows 
“GPS off” in the bottom 
bar. 
This indicates that the GPS is turned off.  Please 
refer to Appendix A.14, GPS Settings (p. A-6) to 
find the GPS power control. 
21 