Instruction manual
HS2 and HS2P 
A soil profile page will be displayed.  This page shows the reference values 
stored for this soil profile and the rod length used.  
To select this soil profile to apply to the current measurement, press   to 
highlight SELECT THIS SOIL and press  .  
New reference values can be set from the soil profile screen.  To do so, 
correctly insert the sensor in the reference soil and select “WET VWC” or 
“DRY VWC” and press  .  
The screen will show MEASURING… while the measurement is taken. After 
3 to 4 seconds, the soil profile page will be shown with the reference value 
changed.  Repeat this process for the remaining wet or dry set point.  
If desired, the soil profile can be cleared by selecting CLEAR SOIL then  .  
The CLEAR SOIL operation cannot be undone. 
The sensor length stored in a soil profile is automatically updated to match the 
sensor used whenever the wet and dry reference values are changed.  If the 
sensor length used to change one of the references is different to that 
previously used for the soil profile, the following screen will be shown. 
NOTE 
11 