Example measurement instructions for CR1000 datalogger, with the sensor
wired to SE channel 2:
Public AirTC
VoltSe (AirTC,1,mV2500,2,0,0,_60Hz,0.09,-60)
Example measurement instruction for CR10X datalogger:
1: Volt (SE) (P1)
1: 1 Reps
2: 5 2500 mV Slow Range
3: 2 SE Channel
4: 1 Loc [ AirTC ]
5: 0.09 Multiplier
6: -60.0 Offset