Product manual
Model HC2S3 Temperature and Relative Humidity Probe
BeginProg
Scan (1,Sec,1,0) 'Run main scan 1 second
Battery (Battery_volts)
PanelTemp (Ptemp_C,250)
'add additional instructions to be executed every 1 second
CallTable (Table1)
NextScan
SlowSequence
Scan (5,Sec,0,0) 'Run slow sequence scan every 5 seconds
PortSet (9,TRUE) 'Turn on HC2S3
Delay (0,3000,mSec) 'Wait 3 seconds for HC2S3 to warm-up
VoltSe (AirTC,1,mV2500,2,0,0,_60Hz,0.1,-40) 'Measure HC2S3 temperature
VoltSe (RH,1,mV2500,1,0,0,_60Hz,0.1,0) 'Measure HC2S3 relative humidity
PortSet (9,FALSE) 'Turn off probe
NextScan
EndProg
5.2 Example Programs using Differential Measurement
Instructions
Temperature and humidity are measured on differential input channels 1 and 2
respectively. The program sets relative humidity = 100 if the measured value
is > 100 but less than 103%. Values > 103% are not set = 100, and indicate a
problem with the sensor or its calibration.
TABLE 5-4. Wiring for
Differential Measurement Examples
Color Description CR1000 CR10(X)
Brown Temperature 1H 1H
Jumper to
Yellow
Temperature Signal
Reference
1L 1L
White Relative Humidity 2H 2H
Yellow Signal Reference 2L 2L
Gray Power Ground G G
Clear Shield
G
Green Power 12V 12V
12