Specifications
Section 3. CR3000 Measurement Details
Reference Temperature Errors Due to Panel Gradient
Chamber Changed from -65 to + 70 deg C
-15
-12
-9
-6
-3
0
0 102030405060708090
Time Minutes
Measured - Actual deg C
-65
-35
-5
25
55
85
Chamber and RefTemp deg C
Chan 1
Chan 2
Chan 3
Chan 4
Chan 6
Chan 7
RefTemp_Avg
Chamber Temp
FIGURE 3.4-2. Panel Temperature Gradients during -65 to 70 °C Change
Reference Temperature Errors Due to Panel Gradient
Chamber Changed from 75 to 25 deg C
-1
0
1
2
3
0 102030405060708090100
Time Minutes
Measured - Actual deg C
0
20
40
60
80
Chamber and RefTemp deg C
Chan 1
Chan 2
Chan 3
Chan 4
Chan 6
Chan 7
RefTemp_Avg
Chamber Temp
FIGURE 3.4-3. Panel Temperature Gradients during 75 to 25 °C Change
3-11