Specifications

CR3000 Overview
Step 3: Output Processing
Step 3 is to define the output processing for the sensor measurements. Click
the Outputs link in the Progress menu at the left.
The Outputs screen has a list of Selected Sensors on the left, and Output
Tables on the right. The default is for two Tables, Table1 and Table2. Both
Tables have a Store Every field and the drop-down list box that are used to set
the interval at which data will be stored.
The objective for this exercise calls for a one-minute output processing. To
remove Table2, Click on the Table2 tab to activate it, and click the Delete
Table button. When the Confirm window appears click Yes.
The Table Name field is the name that will be used for the Table in which the
output will be stored. Change the default Name of Table1 to OneMin, and
change the Store Every Field to 1 minute.
The Selected Sensors list is provided on the left side of the screen. To add a
sensor measurement to the Output Table, highlight a measurement and click
one of the output buttons; e.g., Average. Select the Default, Panel Temp, and
Type T TC sensors and click the Average button to add them to the OneMin
Table.
OV-28