Specifications
Section 7. Measurement Instructions
Parameter
& Data Type
Enter
Integ The time spent on integration in microseconds for each of the channels measured.
Constant Entry
Integration
250
250 µS
_60Hz or 16667
16,667 µS (reject 60 Hz noise)
_50 Hz or 20000
20,000 µS (reject 50 Hz noise)
Mult, Offset
Constant,
Variable,
Array, or
Expression
A multiplier and offset by which to scale the raw results of the measurement.
7.4 Full Bridges
BrFull (Dest, Reps, Range, DiffChan, ExChan, MeasPEx, ExmV, RevEx,
RevDiff, SettlingTime, Integ, Mult, Offset)
X = result w/mult = 1, offset = 0
X
V
V
R
RR
R
RR
x
==
+
−
+
⎛
⎝
⎜
⎞
⎠
⎟
1000 1000
13
34
2
12
This Instruction applies an excitation voltage to a full bridge and makes a
differential voltage measurement of the bridge output. The resulting value
(multiplier = 1, offset = 0) is the measured voltage in millivolts divided by the
excitation voltage in volts (i.e., millivolts per volt).
BrFull6W (Dest, Reps, Range1, Range2, DiffChan, ExChan, MeasPEx,
ExmV, RevEx, RevDiff, SettlingTime, Integ, Mult, Offset)
X = result w/mult = 1, offset = 0
X
V
V
R
RR
R
RR
==
+
−
+
⎛
⎝
⎜
⎞
⎠
⎟
1000 1000
2
1
3
34
2
12
7-8