Operator`s manual

SECTION 7. MEASUREMENT PROGRAMMING EXAMPLES
7-13
This measurement sequence should not be
used to measure temperature on the 207
temperature and RH probe. The longer
excitation/integration time could cause
polarization of the RH element, shifting its
calibration.
The connections for this example are the same
as for Example 7.5, where instruction 11 is used
to measure three 107 Temperature Probes.
PROGRAM
01: P5 AC Half Bridge
01: 3 Reps
02: 2 15 mV slow Range
03: 1 IN Chan
04: 1 Excite all reps w/EXchan 1
05: 4000 mV Excitation
06: 1 Loc [:107 T #1 ]
07: 800 Mult
08: 0 Offset
02: P 55 Polynomial
01: 3 Reps
02: 1 X Loc 107 T #1
03: 1 F(X) Loc [:107 T #1 ]
04: -53.46 C0
05: 90.807 C1
06: -83.257 C2
07: 52.283 C3
08: -16.723 C4
09: 2.211 C5