Operator`s manual
SECTION 13. 21X MEASUREMENTS
13-4
NOTE: Since the peak transient, V
eo
,
causes significant error only if it is several
times larger than the signal, V
so
, error
calculations made in this section
approximate V'
eo
by V
eo
; i.e., V'
eo
≈ V
eo
.
If the input settling time constant, τ, is known, a
quick estimation of the settling error as a
percentage of the maximum error (V
so
for rising,
V'
eo
for decaying) is obtained by knowing how
many time constants (t/τ) are contained in the
450µs 21X input settling interval (t). The
familiar exponential decay relationship is given
in Table 13.3-1 for reference.
TABLE 13.3-1. Exponential Decay, Percent
of Maximum Error vs. Time in Units of τ
ττ
τ
Time % Time %
Constants Max. Error Constants Max. Error
0 100.0 5 0.7
1 36.8 7 0.1
3 5.0 10 0.004
Before proceeding with examples of the effect
of long lead lengths on the measurement, a
discussion on obtaining the source resistance,
R
o
, and lead capacitance, C
w
L, is necessary.
DETERMINING SOURCE RESISTANCE
The source resistance used to estimate the
settling time constant is the resistance the 21X
input "sees" looking out at the sensor. For our
purposes the source resistance can be defined
as the resistance from the 21X input through all
external paths back to the 21X. Figure 13.3-2
shows a typical resistive sensor, (e.g., a
thermistor) configured as a half bridge. Figure
13.3-3 shows Figure 13.3-2 redrawn in terms of
the resistive paths determining the source
resistance Ro, is given by the parallel
resistance of Rs and Rf, as shown in Equation
13.3-8.
FIGURE 13.3-2. Typical Resistive Half Bridge
FIGURE 13.3-3. Source Resistance Model
for Half Bridge Connected to the 21X
R
o
= R
s
R
f
/(R
s
+R
f
) [13.3-8]
If R
f
is much smaller, equal to or much greater
than R
s
, the source resistance can be
approximated by Equations 13.3-9 through
13.3-11, respectively.
R
o
≈ R
f
, R
f
<<R
s
[13.3-9]
R
o
= R
f
/2, R
f
=R
s
[13.3-10]
R
o
≈ R
s
, R
f
>>R
s
[13.3-11]
The source resistance for several Campbell
Scientific sensors are given in column 3 of
Table 13.3-5.
DETERMINING LEAD CAPACITANCE
Wire manufacturers typically provide two
capacitance specifications: 1) the capacitance
between the two leads with the shield floating,
and 2) the capacitance between the two leads
with the shield tied to one lead. Since the input
lead and the shield are tied to ground (often
through a bridge resistor, R
f
) in single-ended
measurements such as Figure 13.3-2, the
second specification is used in determining lead
capacitance. Figure 13.3-4 is a representation
of this capacitance, C
w
, usually specified as
pfd/ft. C
w
is actually the sum of capacitance
between the two conductors and the
capacitance between the top conductor and the
shield. Capacitance for 3 Belden lead wires
used in Campbell Scientific sensors is shown in
column 6 of Table 13.3-2.