Appendix B. SDI-12 Commands/Changing Settings 
B-3 
microwave pulses. Any echo occurring 0.5 m (1.6 ft) short of the distance you 
entered will be considered noise. 
To start False Echo Learn, do the aXSFEL+nnn.nnn! command (where nnn.nnn = 
the actual distance to the water) followed by the aD0! (Send Data) command. 
Table 4-5 in Section 4, Quickstart, shows an example of the command and 
response. 
B.1.2 Set Water Stage 
With the Set Water Stage command, you enter the initial depth of the water, and the 
sensor will automatically measure the distance between the sensor and the water 
surface. The water stage setting and the initial distance measurement are used to 
calculate subsequent water stage measurements. Correct stage measurements 
require that the water stage setting be in the same units as the Units of Measure 
setting. You can find out if the sensor is set to metres or feet by using the aXGU! 
(Get Unit) command. 
To set the water stage, do an aXSS+nnn.nnn! command (where nnn.nnn = the 
initial water depth) followed by the aD0! (Send Data) command. Table 4-6 in 
Section 4, Quickstart, shows an example of the command and response for entering 
this setting. 
B.1.3 Query/Set the Address 
Valid addresses are 0 to 9; A through Z; and a through z. The factory default 
address is set to 0. The address can be verified by sending the sensor the Address 
Query command (see Table B-2). 
Table B-2. SDI-12 Command for Querying the Address 
Initial Command  Response 
?! 
a<cr><If> 
Where a is the current address of the sensor. 
Change the sensor’s address by sending the sensor the aAb! command, where “a” is 
the original address and “b” is the new address. Table B-3 shows an example of the 
command and response for setting the address. 