8.2  Diagnostics and Repair 
The radar sensor is extremely reliable, but problems can occur during operation. 
Most of these problems are caused by the following: 
•  Sensor 
•  Environmental conditions 
•  Power supply 
•  Signal processing 
When you encounter a problem with the radar sensor, check the error messages 
from the aM!, followed by the aD0! command to help evaluate the issue. 
1. During the initial power up or resumption of supply voltage to the 
sensor, some SDI-12 commands, such as the I command, will not 
yield the expected responses. 
2. A typical response to the aD0! command results in a response of 
108003 where approximately 80 seconds is the required time to 
complete the BIST (Built In Self Test) of the instrument. After power 
up is complete, normal SDI-12 communication starts. 
8.2.1  No Measured Value Available — Error E013 
If you are unable to find a measured value, check the following: 
•  Sensor in boot phase 
•  Update the Start False Echo Learn (aXSFEL). 
8.2.2  No Measured Value Available — Error E041, E042, E043 
If you have a hardware error or have defective electronics, try cycling the power to 
the sensor. If the sensor recovers, no further steps are required. If the sensor does 
not recover, do one of the following: 
•  Exchange the electronics module (see Section 8.2.2.1, Exchange Electronics 
Module) 
•  Return the equipment for repair (an RMA number is required) 
8.2.2.1 Exchange Electronics Module 
If you do not have an electronics module onsite, order one from Campbell 
Scientific. 