User manual
CS475, CS476, and CS477 Radar Water Level Sensor 
16 
7.2 Sensor Alignment 
7.2.1 Vertical 
Use a user-supplied bubble level or the #25619 bubble level to make certain the 
antenna horn is aligned within 1° of vertical. The cap needs to be removed when 
using the #25619. If the antenna is not vertical, a trigonometric measurement error 
can occur with respect to the water. The maximum range is reduced because of the 
off-axis return signal. 
7.2.2 Azimuth 
The sensor’s radar beam is polarized so it emits radar waves in an elliptical or 
football shape. You should orient the unit so the lobes are parallel to, and do not 
intersect, the pier when installing on a wall or close to a bridge pier. The radar 
housing has a large hex nut on its mount stem. Two drill marks below the hex nut 
indicate which direction the lobes extend the least. Orient the sensor such that one 
of the marks is aligned towards the wall or pier (see Figure 4-1 and Table 4-4 in 
Section 4, Quickstart). 
7.3  Instrument Housing Adjustment 
After mounting, you can rotate the housing up to 350° to simplify access to the 
conduit entry and terminal compartment. Proceed as follows to rotate the housing 
to the desired position: 
1.  Loosen the set screw on the housing. 
2.  Rotate the housing as desired. 
3.  Tighten the set screw. 
7.4 Wiring 
7.4.1 Datalogger Connection 
As shipped from Campbell Scientific, the sensor is fitted with a cable for connection 
with the datalogger. Appendix A describes replacing this cable. 
Connections to Campbell Scientific dataloggers are given in Table 7-1. The sensor 
should be wired to the channels shown on the wiring diagram created by Short Cut. 
Connect the wires in the order shown in Table 7-1. 
CAUTION










