Specifications
Appendix B. Entering SDI-12 Commands/Changing Settings 
B.1 Changing Settings 
B.1.1 Query/Set the Address 
Valid addresses are 0 to 9; A through Z; and a through z. The factory default 
address is set to 0. The address can be verified by sending the sensor the 
Address Query command (see Table B-2). 
TABLE B-2. SDI-12 Command for Querying the Address 
Initial Command  Response 
?! 
a<cr><If> 
Where a is the current address of the sensor. 
Change the sensor’s address by sending the sensor the aAb! command, where a 
is the original address and b is the new address. Table B-3 shows an example 
of the command and response for setting the address. 
TABLE B-3. Example of Setting Address 
Initial Command  Response 
0A1! 
Where 0 is the original address 
and 1 is the new address 
1<cr><If> 
The new address (1) is set in response. 
B.1.2 Set Units 
The distance measurement can be reported in feet (default) or meters. Change 
the units by first using the aXSU+n! command (where n=1 (feet) or 0 (meters)) 
followed by the aXGU! (Get Units) command. Table B-3 shows an example of 
the command and response for changing this setting. 
TABLE B-4. Example of Setting Units 
Initial Command  Response 
0XSU+0! 
Where the first zero is the sensor 
address and the second zero sets the 
units to meters 
00011<cr><If> 
Where (from left to right), 
0—sensor’s address; 
001—the amount of time (in seconds) 
that you must wait before sending 
another command; 
1—the number of values that will be 
placed in the buffer. 
Subsequent Command  Response 
0XGU! 
Where zero is the sensor address. 
This is the get units command. 
0+0<cr><If> 
Where the first zero is the sensor 
address and the second zero indicates 
that the units are now meters 
B-3 