User Guide
5. To activate the SDI-12 transparent mode on Control Port p, enter
∗8. The TD
datalogger will respond with a “∗” prompt. At the “∗” prompt enter
#. The
TD datalogger will respond with
150000. Finally, enter p (Control Port p)
and press the <enter> key. The TD datalogger will respond with
entering
SDI-12
.
6. To query the CS215 for its current SDI-12 address, enter the command ?!.
The CS215 will respond with the current SDI-12 address.
7. To change the SDI-12 address, enter the command aAb!; where a is the
current address from the above step and b is the new address. The CS215 will
change its address and the datalogger will exit the SDI-12 Transparent Mode.
8. To activate the SDI-12 transparent mode, on Control Port p, enter
∗8. The
TD datalogger will respond with a “∗” prompt. At the “∗” prompt enter
#.
The TD datalogger will respond with
150000. Finally, enter p (Control Port
p) and press the <enter> key. The TD datalogger will respond with
entering SDI-12. Verify the new SDI-12 address by entering the ?!
command. The CS215 will respond with the new address.
9. To exit the SDI-12 Transparent Mode, type in ∗.
FIGURE 2. Screen Capture of SDI-12 Transparent Mode on Edlog Table Based Datalogger using Control
Port 8 and Changing SDI-12 Address from 0 to 1
11